PR231 Multi function calibrator

Product presentation

產品規格

Products Features

1.Temperature difference measurement with accuracy of 0.003 ° C
PR231A calibrator measures the temperature difference between two points in space without other instruments. When the function is used, the four terminals of the source function be used as measurement terminals and the process of the temperature difference data acqusition can be completed within 0.4 Seconds, which effectively improves the accuracy of the measurement. The stability can also be calculated in real time during the test

 

2.Standard temperature measurement

Different from ordinary TC and RTD measurement, the standard temperature measurement may use the certificate value for temperature traceability. The input signals include:STC - > S type, R type, B type, T type.SPRT-> Rtp = 25Ω or Rtp=100Ω.

 

3.Reference Junction Compensation

The reference junction compensation methods of the PR231 series calibrator are very flexible and three methods are available, namely internal, external, and customized. The external reference junction adopts Grade A Pt100, and can input the certificate value for the reference junction data correction. When PR231 series calibrator is combined with PR1501 temperature equalization compensation module, a reference junction  compensation error of less than 0.07°C can be obtained.

 

4.Precision temperature control function 

Using the precision temperature control function, temperature closed-loop control of the constant temperature equipment can be realized instead of the high-precision PID controller. In the case that the constant temperature equipment and grid voltage are satisfactory to the conditions, the temperature fluctuation of the equipment can be better than 0.02°C /10min

(thermostatic bath).

 

5.ρvalue measurement

The PR231 series calibrator can measure the duty factor of the periodic square signal and can be used to verify and calibrate the PID parameters of various digital temperature indicating regulators for time proportional output, and meet the requirements of JJG 617-1996 Verification Regulation of Digital Temperature Indicators and Controllers.

 

6.Thermal calculator

It is used to achieve various conversions between electric and temperature. The conversion support a variety of TCs, RTDs and thermistors.

 

7.Value Settings

The PR231 series calibrator has the most flexible and convenient output value setting method. It is possible to directly set the output value through the numeric keypad, or to increase the increment setting by pressing the direction key. In addition, the device has an editable phase step or slope value setting method. 

 

8.Sinusoidal signal output function

The verification/calibration of some process loggers, especially mechanical recorders, usually involves the operation test. In this case, the user can use the device's sinusoidal signal output function to provide signals for the verified instrument.

The verification/calibration of some process recorders (especially mechanical recorders) usually involves the operation of the test. At this time, the device's sinusoidal signal output function can be used to signal the meter.

 

9.Data logging function

The logging function saves measurement and output data. PR231 series calibrator has powerful record management functions. Up to 32 device numbers can be created. Each device number has 16 logging pages. Each logging page contains four kinds of basic information, namely time, measured value, output value and custom value. Basic information. Users can perform device processing, record deletion, etc. according to their needs.

 

Model selection table

Item PR231A-1 PR231A-2 PR231B-1 PR231B-2
Enhancement model
  •  

  •  

   
Basic model    
  •  

  •  

0.01 grade
  •  

 
  •  

 
0.02 grade      
  •  

 

Basic Parameters

Weight : 990g charging source : 100~240V AC,50~60Hz
Dimension : 225mm*130mm*53mm Working temperature : -10℃~50℃
Cell type : 7.4V 4400mAh, Rechargeable lithium battery Working time : ≥20 hours(24V power off)
Preheating time : 10 minutes after preheating Humidity : 0~80%,Non condensing
Charging time : 5 hours Calibration period : 2 years

 

Performance Index 

1.Basic parameters of measurement:

Function Range Measurement range Resolution 0.01 accuracy 0.02 accuracy Remarks
Voltage 100mV -5mV~120mV 0.1uV 0.005%RD+5uV 0.015%RD+ Input impedance
5uV ≥80mΩ
1V -50mV~1.2V 1uV 0.005%RD+ 0.015%RD+  
10V -0.5V~12V 10uV 0.005%FS 0.005%FS Input impedance
50V -0.5V~50V 0.1mV     ≥1mΩ
Current 50mA -5mA~50 mA 0.1uA 0.005%RD+0.005%FS 0.015%RD+ Internal resistance =10Ω
0.005%FS
Ohm 50Ω 0Ω~50Ω 0.1mΩ 0.005%RD+5mΩ 0.015%RD+ Output 1mA current
5mΩ
500Ω 0Ω~500Ω 1mΩ 0.005%RD+0.005%FS 0.015%RD+
5kΩ 0kΩ~5kΩ 10mΩ 0.005%FS Output 0.1mA current
Thermal couple S、R、B、K、N、J、E、T、EA2、Wre3-25、Wre5-26 0.1℃ / According to ITS-90 scale
Cold end compensation Internal -10℃~60℃ 0.01℃ 0.5℃ 0.5℃
External 0.1℃ 0.1℃
Thermal  Pt10、Pt100、 Pt200、Cu50、Cu100、BA1、BA2、JPt100、 Pt500、 Pt1000 0.01℃ /
resistance
Standard temperature S、R、B、T、SPt25、SPt100 0.01℃ / Need to enter the correction value
ρ-value 50S 0.001%~99.999% 0.00% 0.01% 0.01% Input pulse width amplitude range:1V~50V
Frequency 10Hz 0.001Hz~12Hz 0.001Hz 0.01%FS 0.01%FS
1kHz 0.00001kHz~ 0.01Hz
1.2 kHz
100kHz 0.01kHz~ 10Hz 0.1%FS 0.1%FS
120 kHz
Temperature difference S、R、B、K、N、J、E、T 0.01℃ / Need to enter the correction value
SPt25、SPt100 0.001℃

 

2.Basic parameters of the output function:

Function Range Measurement
range
Resolution 0.01 accuracy 0.02 accuracy Remarks
Voltage 100mV -20mV~120mV 1uV 0.005%RD+5uV 0.015%RD+5uV Maximum load current =2.5mA
1V -0.2mV~1.2V 10uV 0.005%RD+0.005%FS 0.015%RD+0.005%FS
10V -2V~12V 0.1mV
Current 30mA -5mA~30 mA 1uA 0.005%RD+0.005%FS 0.015%RD+0.005%FS Maximum load voltage =24V
Ohm 50Ω 0Ω~50Ω 0.1mΩ / According to ITS-90 scale
500Ω 0Ω~500Ω 1mΩ
5kΩ 0kΩ~5kΩ 10mΩ
Thermal couple S、R、B、K、N、J、E、T、EA2、Wre3-25、Wre5-26 0.1℃ /
Thermal Pt10、Pt100、 Pt200、Cu50、Cu100、BA1、BA2、JPt100、 Pt500、 Pt1000 0.01℃ /
resistance
Frequency 10Hz 0.001Hz~ 0.001Hz 0.01%FS 0.01%FS Maximum load current =2.5mA
/ Pulse 12 Hz
  1kHz 0.00001kHz~  
  1.2 kHz 0.01Hz
  100kHz 0.01kHz~120 kHz 10Hz 0.1%FS 0.1%FS
Precision temperature control S、R、B、K、N、J、E、T 0.01℃ /
Pt100
24V output Maximum voltage error: 0.3V   Ripple noise:35mVp-p(20MHz bandwidth)
Maximum load current:70mA  Load regulation: 0.5%(10% -100% load change)

 

產品介紹

PR231 Precision Multifunction Calibrator

The PR231 series products have excellent performance indicators, numerous practical functions and powerful human-computer interaction capabilities. The product covers two accuracy levels of 0.01 and 0.02. The measurement and output circuits are completely isolated. In addition to the common functions of the common two-channel calibrator, it also has the functions of p-value measurement and standard temperature measurement.